VGC50x Vacuum Gauge Controller

VGC50x Vacuum Gauge Controller


Specifications
Details

Compatible to the wide range of INFICON active gauges, the new VGC50x series of active gauge controllers are able to monitor and data log the entire pressure range from 10⁻¹⁰ to 1500 mbar (10⁻¹⁰ to 1125 Torr) and the set point status.

 

Advantages

  • Simple operation with dot matrix menu guided display for parameter, sensor or general settings
  • Very bright and clear Display for long distance Instrument read- out
  • Bar graph display with setpoints's or pressure vs. time display
  • Data log and parameter log function with USB port (rear side) and USB stick (front side)
  • Ethernet interface
  • Firmware upgrade available online or with USB stick
  • Two free definable setpoints per channel with adjustable hysteresis
  • High Resolution - 16 bit A/D converter
  • Automatic identification of the INFICON active gauges
  • Programmable 0 to 10 V Chart Recorder Output with logarithmic/ linear characteristics for single gauge or gauge combination (only VGC502 and VGC503)

Specifications

VGC501
VGC502
VGC503
Type   VGC501
Measurement channels   1
Display   LCD, LED backlight
Rate 1/s 10
Connectable gauges with display range
CDG (A/D) Torr 1×10-3×F.S. …1×F.S.
PCG mbarTorr 5×10-4…1500
PSG mbarTorr 5×10-4…1000
MPG mbarTorr 1×10-9…1000
PEG mbarTorr 1×10-9…1×10-2
MAG mbarTorr 1×10-9…1×10-2
BCG mbarTorr 5×10-10…1500
BPG mbarTorr 5×10-10…1000
HPG mbarTorr 2×10-6…1000
Measurement unit (selectable)   mbar, Torr, Pa, hPa, Micron, V
Setpoints
Setpoint relays Setpoint relays 2
Channel assignment Channel assignment 1
Adjustment range Adjustment range sensor dependent
Hysteresis Hysteresis adjustable
Relay contact Relay contact floating changeover contact
Contact rating V (ac) • AV (dc) • A 30 / 1
Connector Connector D-sub, 15-pin, male
Analog output
Range V 0 ... 10.3, sensor analog output signal
Analog output Analog output 1
Programmable analog output Programmable analog output -
Connector Connector D-Sub, 15-pin, male
Interface (digital)
Connector Connector USB slave, master and Ethernet, USB Typ A (stick), USB Type B, FCC68/RJ45
Power
Supply V (ac) 100 ... 240
Frequency Hz 50 ... 60
Consumption W ≤45
Operation temperature (ambiance) °C +5 ... +50
 


View more about VGC50x Vacuum Gauge Controller on main site